Catherine Y. Spong
In the beginning of the 19th century, reports on the presence of fetal heart tones were published, and nearly 150 years later, continuous fetal heart rate (FHR) monitoring became a reality. By 1998, electronic fetal monitoring was used in 84% of all U.S. births, regardless of whether the primary caregiver was a physician or a midwife. With the advent of these technologies, fetal monitoring is implemented in nearly all pregnancies, either in the antepartum or intrapartum period. The challenge of fetal surveillance is to identify those fetuses whose physiological defense mechanisms are compromised, in order to be able to act before decompensation has occurred. The goal is to prevent fetal and neonatal morbidity, and especially mortality.
WHY PERFORM FETAL MONITORING?
Since its inception, the primary objective of FHR monitoring has been to identify the fetus in distress so that measures might be taken in time to avert permanent fetal damage or death. However, a clear consensus regarding the definition of “fetal distress” has not been established. It has been described as “a condition in which fetal physiology is so altered as to make death or permanent injury a probability within a relatively short period of time,” and is usually considered to denote disruption of normal fetal oxygenation, ranging from mild hypoxia to profound fetal asphyxia. The term hypoxia refers to the reduction of tissue oxygen supply below physiologic levels. Asphyxia, derived from the Greek word meaning “a stopping of the pulse,” implies a combination of hypoxia and metabolic acidosis. Historically, the clinical diagnosis of birth asphyxia has been based on findings such as meconium-stained amniotic fluid, abnormal FHR patterns, low Apgar scores, abnormal blood gases, and neonatal neurologic abnormalities. When present together, these findings are highly suggestive of a recent asphyxial insult. Isolated abnormalities, however, correlate poorly with birth-related asphyxia and subsequent neurologic impairment. In 2002, the American College of Obstetricians and Gynecologists Task Force on Neonatal Encephalopathy and Cerebral Palsy stated the criteria to define an acute intrapartum event sufficient to cause cerebral palsy.
Essential criteria (must meet all four):
1. Evidence of a metabolic acidosis in fetal umbilical cord arterial blood obtained at delivery (pH <7 and base deficit ≥12 mmol/L).
2. Early onset of severe or moderate neonatal encephalopathy in infants born at 34 or more weeks of gestation.
3. Cerebral palsy of the spastic quadriplegic or dyskinetic type.
4. Exclusion of other identifiable etiologies such as trauma, coagulation disorders, infectious conditions, or genetic disorders.
Criteria that collectively suggest an intrapartum timing (within close proximity to labor and delivery, eg., 0–48 hours) but are nonspecific to asphyxial insults:
1. A sentinel (signal) hypoxic event occurring immediately before or during labor.
2. A sudden and sustained fetal bradychardia or the absence of fetal heart rate variability in the presence of persistent, late, or variable decelerations, usually after a hypoxic sentinel event when the pattern was previously normal.
3. Apgar scores of 0–3 beyond 5 minutes.
4. Onset of multisystem involvement within 72 hours of birth.
5. Early imaging study showing evidence of acute nonfocal cerebral abnormality.
At the cellular level, asphyxia triggers a cascade of events, including membrane depolarization, disruption of energy metabolism, altered neurotransmission, ion shifts, protease activation, free radical production, and phospholipid degradation. Profound and prolonged asphyxia may result in cell death and, eventually, death of the organism. Sublethal asphyxia may lead to multi-organ system dysfunction. Severe asphyxial brain injury may lead to long-term neurodevelopmental impairment.
Cerebral palsy (CP) is a major disorder of neurodevelopment, defined as “a chronic disability, characterized by aberrant control of movement and posture, appearing early in life and not the result of recognized progressive disease.” It may be accompanied by mental retardation (41%), seizures (23%), or cortical visual impairment. The insult responsible for the development of CP may occur at any time during the prenatal, perinatal, or postnatal periods. Unlike other major neurodevelopmental disorders, the relationship between CP and abnormal or difficult birth has long been recognized, publicly dating back to a treatise presented by William John Little in 1862. In 1943, Windle demonstrated clinical and histopathologic evidence of neural damage in experimentally asphyxiated fetal guinea pigs. He later reported the effects of prolonged anoxia on fetal rhesus monkeys. Total anoxia for less than 8 minutes did not produce consistent injury, whereas anoxia for more than 10 minutes invariably resulted in neuropathology. There were no survivors beyond 20 to 25 minutes of anoxia. The pattern of injury produced by prolonged anoxia, however, did not correlate with the cerebral injury, mental retardation, and spasticity seen in CP. Later work demonstrated that prolonged partial asphyxia in monkeys produced acidosis, late FHR decelerations, and neuropathologic defects consistent with the findings in the common forms of CP. In addition to lesions in the thalamus and basal ganglia, prolonged partial asphyxia caused generalized cerebral necrosis or focal necrosis in the parasagittal regions and the border zones between the parietal and occipital lobes.
Although early studies created and fostered the assumption that birth-related asphyxia was the primary cause of CP, recent evidence challenges this assumption. In 1986, Nelson and Ellenberg reported a multivariate analysis of risk in 189 cases of CP. After accounting for major congenital malformations, low birth weight, microcephaly, and alternative explanations for the disorder, they were able to attribute only 9% of CP cases to birth asphyxia. Others have reached similar conclusions. Although birth asphyxia nearly tripled the odds of developing CP, only 8.2% of CP cases were potentially attributable to birth asphyxia.
As early as the 19th century, researchers using auscultation recognized that certain FHR patterns were associated with adverse perinatal outcome. The introduction of direct electronic fetal monitoring and fetal scalp blood sampling in the 1960s provided tools for evaluating the fetus. FHR decelerations have been found to be correlated with fetal acidosis. Fetuses with no decelerations, early decelerations, or mild variable decelerations had average scalp pH values greater than or equal to 7.29, whereas those with severe variable or late decelerations had pH values less than or equal to 7.15. In addition, FHR variability was found to be correlated with scalp pH values as fetuses with normal FHR variability had higher scalp pH values than those with decreased variability. The absence of FHR accelerations was correlated with poor perinatal outcome and the presence of FHR accelerations has been shown to predict normal scalp pH values. With the development of indirect monitoring techniques, the experience derived from direct intrapartum monitoring became applicable to the antepartum period, leading to the development of antepartum testing.
Antepartum fetal monitoring has the goal of identifying the fetus at risk, allowing sufficient time to intervene before permanent injury or death occur. Intrapartum fetal monitoring should be able to identify three groups of fetuses:
· the fetus that is not affected by labor
· the fetus that is negatively affected by labor but has enough reserve to compensate fully and is in no immediate danger
· the fetus that is negatively affected and lacks the reserve to compensate, thus uses its key resources to survive and is in danger for morbidity/mortality.
It is the third group that would most benefit from intervention.
WHO SHOULD BE MONITORED?
Antepartum fetal monitoring is typically offered to patients at increased risk of fetal or neonatal morbidity/mortality. These include maternal medical complications (e.g., diabetes mellitus, chronic hypertension, systemic lupus erythematosis), fetal conditions (Rh isoimmunization, fetal growth restriction), and pregnancy complications (multiple gestation, oligohydramnios). Overall, the studies supporting the methods of testing, the timing, and initiation are extremely varied; thus absolute guidelines based on scientific evidence cannot be established. However, recommended gestational ages for initiation of testing, timing, and methods for specific maternal conditions, and the supportive evidence can be found in Table 9.1.
|
|
|
TABLE 9.1. General guidelines for initiation of antenatal testing |
Intrapartum fetal monitoring has remained controversial since its inception. The FHR many be evaluated by auscultation or by electronic monitoring. Auscultation is typically performed with a DeLee stethoscope or Doppler ultrasound. Electronic monitoring can be either performed externally or internally, external monitors use a Doppler device with computerized logic to interpret and count the signals. Internal monitoring uses a fetal electrode that records the fetal electrocardiogram (ECG). Well-controlled studies have shown the equivalence of intermittent auscultation to continuous fetal heart monitoring when auscultation was performed at specific intervals with a 1:1 nurse-to-patient ratio. The intensity of monitoring is based on risk factors, with more intensive surveillance required for high-risk pregnancies.
WHAT CAN WE MONITOR?
Fetal Heart Rate
The FHR can be monitored and recorded indirectly through the use of an ultrasound transducer or directly via a subcutaneous ECG electrode placed on the fetus (Fig. 9.1). The indirect method can be used throughout pregnancy and has no contraindications. Using the indirect method, ultrasound waves originating from the transducer penetrate the tissues and are reflected by tissue interfaces. Waves reflected from the moving structures of the fetal heart return to the transducer and are translated into electrical signals. In the direct method, the subcutaneously placed ECG electrode detects electrical impulses originating in the fetal heart. Amplified signals are processed by a cardiotachometer, comparing each incoming QRS complex to the one immediately preceding it. The interval between the two complexes is used to calculate a heart rate. The process is repeated with each cardiac cycle yielding a graphic beat-to beat display of the FHR. The direct method requires rupture of the fetal membranes for placement of the ECG electrode. In addition, in light of the potential risk of infection, the electrode for direct assessment should only be used when the benefits outweigh this risk.
|
|
|
FIG. 9.1. Direct and indirect fetal monitoring. A: Direct method. Recordings are made from a fetal electrocardiograph electrode applied directly to the fetus. A transcervical intrauterine pressure catheter is used to monitor the strength of uterine contractions. B: Indirect method. The fetal heart rate is derived from a Doppler ultrasound transducer applied to the maternal abdominal wall. A pressure sensor (tocodynamometer) detects uterine contractions. (Courtesy of Richard H. Paul.) |
Uterine Activity
Similar to FHR, the detection and measurement of uterine activity can be performed indirectly or directly (see Fig. 9.1). Indirect assessment of uterine activity is performed with a pressure transducer (tocodynamometer) applied tightly to the maternal abdomen over the uterine fundus. Uterine contractions exert pressure on the abdominal wall that is transmitted to the tocodynamometer. Changes in pressure are converted into signals and plotted on the uterine activity graph. The indirect method is noninvasive and can be performed at any time during pregnancy. The limitations of the indirect method are that the readout can only be used to determine contraction frequency, not strength of the contraction (a tightly fit belt will record larger contractions than a loosely fitting or misplaced belt) and the ability to detect contractions in extremely obese women may be difficult. Direct assessment of uterine activity employs a thin, flexible intrauterine pressure catheter (IUPC) placed transcervically into the amniotic cavity. Intrauterine pressure is transmitted from the amniotic fluid through the fluid-filled IUPC to a pressure transducer. The transducer converts pressure measurements into electrical signals, and continuous pressure readings are displayed on the uterine activity graph. The direct assessment is invasive and requires ruptured membranes. Aside from some research indications, the IUPC should only be used in patients when the benefits outweigh the risk of infection. The direct method allows the readout of both the frequency and the strength of the uterine contractions. This can be especially useful in the evaluation and assessment of patients with prolonged labor.
Fetal State (Tone/Breathing/Movements)
Using real-time ultrasound, the state of the fetus can be evaluated. Typically included in this evaluation are assessments of fetal tone, movements, and breathing. Fetal voiding and swallowing can also be evaluated. Specific assessments of fetal tone, breathing, and movements are discussed later in the chapter “Biophysical Profile.”
Amniotic Fluid Volume
The volume of amniotic fluid is a measure of fetal well-being. By the second trimester the predominant source of amniotic fluid is fetal urine. The level of amniotic fluid is thought to represent “long-term” fetal well-being. A compromised fetus will preferentially shunt blood to the major organs (central nervous system, adrenals) and away from others, such as the kidney. Decreased fetal renal perfusion results in a decrease in fetal renal function and subsequent oligohydramnios. The amniotic fluid can be assessed ultrasonongraphically. There are a number of methods to quantitate the volume, including the amniotic fluid index, single deepest pocket, two-dimensional pocket, and a subjective assessment.
HOW DO WE MONITOR?
Equipment
The fetal monitor tracing is a continuous paper strip composed of two Cartesian graphs. The FHR tracing is displayed on the upper graph, with time on the x-axis and heart rate on the y-axis (range 30 to 240 beats per minute). Uterine activity is displayed on the lower graph, with time on the x-axis and pressure on the y-axis (range 0 to 100 mm Hg). Heart rate and uterine activity are plotted separately on the heat-sensitive paper by two thermal pens. On both grids, fine vertical lines represent 10-second intervals, and heavy lines denote 1-minute intervals. In the United States, the standard paper speed is 3 cm per minute.
INTERPRETATION OF THE FETAL MONITOR
Analysis of the fetal monitor strip requires a systematic approach. First, the FHR is analyzed with respect to (a) the baseline, (b) variability, and (c) periodic patterns, including FHR accelerations and decelerations (Table 9.2). Uterine activity is evaluated with attention to the frequency, duration, and strength of contractions, as well as the baseline uterine tone between contractions.
|
|
|
TABLE 9.2. Fetal monitor interpretation |
FETAL HEART RATE INTERPRETATION
Baseline Fetal Heart Rate
The normal FHR baseline ranges from 120 to 160 beats per minute. Early in pregnancy, it is closer to 160 beats per minute, declining as gestational age advances. Likewise, the FHR may decrease gradually toward 120 beats per minute during the course of labor. An FHR baseline below 120 beats per minute is termed bradycardia, and a rate in excess of 160 beats per minute is termed tachycardia. Abnormalities in the FHR baseline may have very different causes and consequences. It is important therefore to characterize the underlying etiology as accurately as possible and to institute appropriate therapy at the earliest possible time.
Bradycardia
Bradycardia is defined as an abnormally low baseline FHR (< 120 beats per minute) and must be differentiated from the episodic FHR changes characteristic of decelerations. Although FHR decelerations are very common, true fetal bradycardia is not. A bradycardic FHR baseline between 100 and 120 beats per minute observed in association with otherwise reassuring FHR patterns probably represents a normal variant. Rarely, fetal bradycardia may be seen in association with maternal β-blocker therapy, hypothermia, hypoglycemia, hypothyroidism, or fetal cardiac conduction defects (congenital atrioventricular block). Documentation of fetal heart block should prompt a search for structural fetal cardiac abnormalities, which may be present in 20% of cases. Other causes of heart block include viral infections (e.g., cytomegalovirus) and damage to the cardiac conduction system by transplacental passage of maternal anti-Ro (anti–SS-A) antibodies. Most congenital causes of fetal bradycardia do not present as abrupt changes in the FHR and rarely require emergency intervention. Any abrupt decline in the FHR below 120 beats per minute more likely represents a deceleration than a change in the baseline, and should be considered pathologic until proven otherwise.
Tachycardia
Fetal tachycardia has many possible etiologies. Most often, it is the result of decreased vagal or increased sympathetic outflow, associated with fever, infection, fetal anemia, or fetal hypoxia. Other causes include maternal hyperthyroidism, fetal tachyarrhythmias (e.g., paroxysmal supraventricular tachycardia, atrial fibrillation, atrial flutter, and ventricular tachyarrhythmias), and medications including sympathomimetics (e.g., ritodrine, terbutaline) and parasympatholytics (e.g., atropine, phenothiazines).
Sinusoidal Pattern
The sinusoidal FHR pattern is an uncommon FHR baseline abnormality. It has the appearance of a smooth sine wave with an amplitude of 5 to 15 beats per minute and a frequency of 2 to 5 cycles per minute. There is little beat-to-beat variability, and accelerations are absent. Although the pathophysiologic mechanism is unclear, this pattern classically is associated with hypoxia and severe fetal anemia. Additionally, it has been reported in association with fetomaternal hemorrhage, chorioamnionitis, fetal sepsis, and administration of narcotic analgesics. A persistent sinusoidal pattern that is not attributable to medications is a concerning finding and demands immediate evaluation.
Fetal Heart Rate Variability
Variability in the FHR results from constant interplay between the sympathetic and parasympathetic arms of the fetal autonomic nervous system. Modulation of vagal tone occurs in response to changes in blood pressure detected by baroreceptors in the fetal aortic arch. Oxygen and carbon dioxide fluctuations, detected by chemoreceptors in the carotid bodies, similarly influence vagal outflow. In the absence of stress, sympathetic outflow is thought to be relatively tonic. Continual adjustments in vagal tone are manifested in the FHR tracing as “short-term” (beat-to-beat) variability superimposed on broader, cyclical fluctuations of 3 to 5 cycles per minute, referred to as “long-term” variability. In clinical use, the term FHR variability refers to the composite of short-term and long-term variability and is quantitated by measuring the difference between the peaks and troughs of the long-term fluctuations (Fig. 9.2). FHR variability is considered normal or average when both short-term and long-term variability are present, and the difference between the peaks and troughs of the long-term fluctuations is 6 to 25 beats per minute. Average variability reflects a nonacidotic vagal connection between the fetal central nervous system (CNS) and the cardiac conduction system. Increased variability (more than 25 beats per minute), or saltatory FHR pattern, is uncommon and most often represents an exuberant autonomic response of a normal fetus. On occasion, it may reflect increased catecholamine release in the early stages of fetal hypoxia. Careful evaluation of the associated FHR findings should help to clarify such cases. Decreased (3 to 5 beats per minute) or absent (0 to 2 beats per minute) FHR variability reflects diminished fetal CNS activity, usually attributable to fetal sleep cycles or to medications administered to the mother (e.g., analgesics, magnesium sulfate, benzodiazepines, phenothiazines, atropine). Persistently decreased variability, however, may signal fetal acidosis. This is particularly true in the presence of other FHR findings suggestive of hypoxia, including tachycardia, loss of reactivity, or repetitive decelerations.
|
|
|
FIG. 9.2. Fetal heart rate variability. A: Short-term variability absent, long-term variability absent—abnormal. B: Short-term variability present, long-term variability absent—abnormal. C: Short-term variability absent, long-term variability present—abnormal. D: Short-term variability present, long-term variability present—normal. |
Periodic Patterns
The FHR baseline is frequently interrupted by accelerations or decelerations in rate. These periodic patterns have important clinical implications regarding the well-being of the fetus.
Accelerations
Accelerations in the FHR occur with 90% of fetal movements as early as the second trimester, probably as a result of increased catecholamine release and decreased vagal stimulation of the heart (Fig. 9.3). By 32 weeks gestation, nearly all normal fetuses will have 15 to 40 spontaneous accelerations per hour, reflecting normal oxygenation of the CNS–cardiac axis. The frequency and amplitude of accelerations may be diminished by fetal sleep states, medications (narcotics, magnesium sulfate, atropine), prematurity, or fetal acidosis. Often, fetal scalp stimulation or vibroacoustic stimulation will provoke fetal movement and FHR accelerations. If these measures fail to induce FHR accelerations, hypoxia should be suspected, particularly if other FHR characteristics are not reassuring.
|
|
|
FIG. 9.3. Fetal heart rate accelerations. |
Decelerations
Decelerations in the FHR are most commonly encountered during the intrapartum period. They are divided into three categories: early, variable, and late decelerations (Fig. 9.4). Classification is based on the characteristic appearance of the deceleration and its temporal relationship to the onset of a uterine contraction.
|
|
|
FIG. 9.4. Fetal heart rate decelerations. |
Early Decelerations
Early decelerations are typically uniform, shallow dips in the FHR (rarely below 100 beats per minute) that mirror uterine contractions, beginning at the onset of the contraction and ending when the contraction ends. They are thought to result from fetal head compression, transient elevation of intracranial pressure, and reflex augmentation of vagal tone. Early decelerations classically appear during labor when the cervix is dilated 4 to 6 cm. Perinatal outcome is not adversely affected by these decelerations, and they are considered clinically benign.
Variable Decelerations and Prolonged Decelerations
Variable decelerations result from umbilical cord compression. They are abrupt and angular in appearance and have a variable temporal relationship to uterine contractions. As the umbilical cord is compressed, the thin-walled, compliant umbilical vein is the first vessel occluded, resulting in decreased fetal venous return, relative hypovolemia, and a reflex increase in the FHR. This observation often is termed a “shoulder.” Further compression of the umbilical cord leads to occlusion of the umbilical arteries, removing the low-resistance placenta from the circuit and dramatically increasing fetal peripheral resistance. This leads to elevation of the fetal blood pressure and a baroreceptor-mediated slowing of the FHR in an attempt to return the blood pressure to normal. Maximum vagal tone may result in a junctional or idioventricular escape rhythm that appears as a relatively stable rate of 60 to 70 beats per minute at the nadir of the deceleration. As cord compression is relieved, this sequence of events occurs in reverse, at times resulting in transient tachycardia or “overshoot” at the end of the deceleration. Variable decelerations are classified as:
· Mild—duration less than 30 seconds, depth of deceleration less than 70 beats per minute
· Moderate—duration 30 to 60 seconds, depth less than 70 beats per minute
· Severe—duration greater than 60 seconds, depth less than 70 beats per minute.
Isolated, infrequent variable decelerations have little clinical significance. Repetitive severe variables, however, may not allow sufficient fetal recovery between decelerations, resulting in persistent hypoxemia, hypercapnia, and respiratory acidosis. Prolonged tissue hypoperfusion may lead to metabolic acidosis and, ultimately, fetal death. In animal models, Clapp reported that frequent episodes of hypoxemic stress, produced by intermittent umbilical cord occlusion over a period of hours, produced fetal injury even in the absence of acidosis. When repetitive, severe variable decelerations are present, prolapse of the umbilical cord must be excluded. Other causes include nuchal cord, true knot in the cord, uterine rupture, placental abruption, uterine hypertonus, and tachysystole. Occasionally, variable decelerations fail to return promptly to the baseline and may more accurately be termed prolonged decelerations. Prolonged decelerations usually result from umbilical cord compression (cord prolapse, nuchal cord) or other acute interruption of uteroplacental transfer of oxygen (tetanic contraction, uterine rupture, maternal hypotension, maternal apnea, placental abruption).
Late Decelerations
Late decelerations reflect inadequate uteroplacental transfer of oxygen during contractions. Typically, they are smooth, uniform decelerations that begin after the onset of a contraction and end after the contraction stops. During uterine contractions, decreased maternal perfusion of the uteroplacental unit causes a decline in fetal PO2. Fetal PO2 levels below a critical threshold of 15 to 18 mm Hg trigger a complex chemoreceptor- and baroreceptor-mediated reflex. Initially, centralization of blood volume (favoring perfusion of the brain, heart, and adrenals) occurs via vasoconstriction in the vascular beds of the limbs and gut. The resulting increase in peripheral resistance provokes a reflex deceleration in the FHR. Isolated late decelerations within an otherwise normal tracing have little clinical significance. However, repetitive hypoxemia and centralization of blood volume, as evidenced by repetitive late decelerations, may force hypoperfused tissues to convert from aerobic to anaerobic metabolism. Organic acid by-products of anaerobic metabolism (pyruvate, lactate) diffuse slowly across the placenta and may accumulate in the fetus, leading to metabolic acidosis, asphyxia, and possibly death. Late decelerations may be caused by any factor that (a) reduces the normal placental transfer of oxygen or (b) increases the fetal oxygen demand beyond the available supply. Such factors include uterine hypertonus or tachysystole (oxytocin, prostaglandins, uterine rupture, placental abruption), maternal hypertension (chronic hypertension, preeclampsia, collagen vascular disease, renal disease, diabetes), suboptimal maternal cardiac output (cardiac disease, hypovolemia, supine hypotension, sympathetic blockade from regional anesthesia, sepsis), maternal hypoxia (apnea, cardiac disease, pulmonary disease), reduced oxygen-carrying capacity of maternal blood (anemia, hemoglobinopathy), and fever (increased fetal metabolism and increased oxygen consumption).
Uterine Activity
Uterine activity is evaluated with attention to the frequency, duration, and strength of contractions, as well as the baseline uterine tone between contractions. The frequency is assessed by counting the number of contractions as recorded over a period of time. The normal frequency of uterine contractions during labor is every 2 to 3 minutes; however, cervical change may result from contractions occurring less frequently. In most cases, five to seven uterine contractions within a 15-minute period reflect adequate uterine activity. Increased contraction frequency is termed uterine tachysystole and is defined by six or more uterine contractions within a 10-minute window for two consecutive windows. Tachysystole may be seen in cases of placental abruption or hyperstimulation of the uterus by oxytocin or prostaglandin cervical-ripening agents. The duration is assessed by calculating the length of the contraction, the fine vertical lines represent 10-second intervals, and heavy lines denote 1-minute intervals (run at 3 cm per minute). Normal uterine contractions last approximately 40 to 60 seconds. Prolonged contractions may result from “coupling” of contractions, uterine hyperstimulation, or acute complications such as placental abruption or uterine rupture. The strength of the contractions can only be assessed with the direct IUPC. The normal baseline uterine pressure between contractions is approximately 10 mm Hg. Abnormally high baseline pressures in excess of 20 mm Hg may result from hyperstimulation or occasionally from overdistention of the uterus by excessive amnioinfusion, polyhydramnios, or fetal macrosomia. During contractions, normal uterine pressure ranges from 30 to 80 mm Hg, although pressures in excess of 80 mm Hg may be observed during the second stage of labor. A calculation of the strength of the contraction is based on the sum of contraction pressures (peak minus baseline) and is expressed in Montevideo units. For example, four contractions in a 10-minute window, each 50 mm Hg above the baseline, would yield 200 Montevideo units. Montevideo units in excess of 180 to 200 usually are considered to reflect adequate uterine activity to effect cervical change in labor.
Antepartum Testing/Fetal Monitoring
Early in its development, electronic FHR monitoring required direct access to the fetus and was limited to the intrapartum period. Later, Doppler ultrasound technology made it possible to monitor the FHR before labor. The experience gained from intrapartum monitoring was applied to the antepartum period and led to the development of antepartum testing. The goals of antepartum testing are (a) to identify fetuses in jeopardy so that permanent injury or death might be prevented and (b) to identify healthy fetuses so that unnecessary intervention might be avoided. The key measure of the effectiveness of an antepartum test is the false-negative rate, defined as the incidence of fetal death within 1 week of a normal antepartum test. Reported false-negative rates range from 0.4 to 1.9 per 1000 with current testing methods. Another important measure is the false-positive rate. A false-positive test may be defined as an abnormal test that prompts delivery but is not associated with evidence of acute fetal compromise (e.g., meconium-stained amniotic fluid, intrapartum fetal distress, low Apgar scores) or chronic fetal compromise (e.g., fetal growth restriction). False-positive rates range from 30% to 90% with current testing methods.
Antepartum testing is used primarily in patients who are considered to be at increased risk for fetal hypoxia or asphyxia secondary to suboptimal uteroplacental transfer of oxygen. It is expected that the use of antepartum testing in these high-risk patients will reduce their risk of fetal/neonatal morbidity/mortality to the level of the low-risk patient. The optimal gestational age at which to begin antepartum testing is not known and varies according to the underlying condition. However, for most medical indications, testing is initiated by 32 to 34 weeks. In view of the high false-positive rates of most testing protocols, earlier initiation of testing should be expected to increase the incidence of unnecessary intervention and iatrogenic prematurity, with its attendant complications.
Fetal Movement Counts
Maternal perception of normal fetal movement has long been recognized as a reliable indicator of fetal well-being. Conversely, prolonged absence of fetal movement may signal fetal death. Cessation of fetal movement in response to hypoxia has been demonstrated in animal studies; however, controlled data in human fetuses are lacking. Nevertheless, any acute decrement in the number or strength of fetal movements should raise the suspicion of fetal compromise and should prompt further evaluation. Many clinicians recommend routine fetal movement counting, particularly in patients who are considered high-risk. A common approach is to recommend daily counting of fetal movements for 1 hour. Ten fetal movements in a 1-hour period are considered reassuring. If fewer than 10 movements are appreciated, counting is continued for another hour. Fewer than 10 movements in a 2-hour period should alert the patient to contact her physician for further evaluation. Another protocol calls for movement counting two to three times daily for 30 minutes. With this approach, further evaluation is recommended if there are fewer than four strong movements in a 30-minute period.
Although controlled trials in low-risk patients have not demonstrated a significant benefit of formalized movement counting over routine questioning during prenatal visits, evidence from one study using nonconcurrent controls suggests a lower rate of fetal death and a higher incidence of intervention for fetal distress in patients using a formalized protocol of fetal movement counting. Fetal movement counting is an inexpensive method of involving the patient in her own care and may be a valuable adjunct to routine prenatal care, regardless of risk category. Some fetal monitors have the capability to display a Doppler-detected fetal movement profile. This technology may aid the clinician in determining the true baseline FHR and has been shown to reduce the incidence of nonreactive nonstress tests by half.
Nonstress Test
FHR accelerations occurring in association with fetal movements form the basis of the nonstress test (NST). Although many criteria have been reported, a normal or “reactive” NST usually is defined by two accelerations in a 10- to 20-minute period, each lasting at least 15 seconds and peaking at least 15 beats per minute above the baseline. In most institutions, the test is repeated once or twice weekly. Boehm and colleagues (1986) reported that the latter approach yielded a three-fold reduction in the incidence of fetal death, although the difference was not statistically significant. Freeman and co-workers (1982) reported a false-negative rate of 1.9 per 1000 among 1542 women tested weekly with the NST. Manning and associates (1983) reported an average false-negative rate of 6.4 per 1000 among nine large clinical trials using the NST as the primary method of surveillance. Assessment of FHR characteristics other than reactivity (baseline rate, variability, decelerations) may improve the sensitivity of the test. The findings from an NST must be interpreted in the context of the pregnancy. Preterm infants commonly do not meet the 15 beats per 15 seconds criterion for reactivity. Once the fetus is 34 weeks, the tracing should be reactive. However, once a fetus demonstrates FHR reactivity, even if earlier than 34 weeks gestation, it should continue to have reactive tests. Reported false-positive rates of the NST vary widely, with an average rate of approximately 50%. When performed twice weekly and interpreted in the context of associated FHR patterns, the NST alone appears to be an acceptable, though not optimal, method of antepartum testing. Advantages include ease of use and interpretation, low cost, and minimal time requirement. The chief disadvantages include a high false-positive rate and a higher false-negative rate than achieved with other methods. Commonly the NST is used as a screening test, and if abnormal, a backup test such as the contraction stress test or biophysical profile is used.
Contraction Stress Test and Oxytocin Challenge Test
The first antepartum testing technique, the contraction stress test (CST) or oxytocin challenge test (OCT), arose from intrapartum observations linking late FHR decelerations with poor perinatal outcome. The test sought to identify uteroplacental insufficiency by demonstrating late decelerations in fetuses exposed to the stress of spontaneous (CST) or induced (OCT) uterine contractions. Late decelerations occurring during uterine contractions were associated with increased rates of fetal death, growth retardation, and neonatal depression. The CST is performed weekly and is considered negative if there are at least three uterine contractions in a 10-minute period with no late decelerations on the tracing. Failure to produce three contractions within a 10-minute window or inability to trace the FHR results in an unsatisfactory test. Prolonged decelerations, variable decelerations, or late decelerations occurring with fewer than half of the contractions constitute a suspicious or equivocal test. Unsatisfactory, suspicious, or equivocal tests require repeat testing the following day. The CST or OCT is considered positive when at least half of the contractions during a 10-minute window are associated with late decelerations. Freeman and colleagues (1982) tested more than 4600 women with the CST and reported a false-negative rate of 0.4 per 1000. When the last test before delivery was a reactive, negative CST, the perinatal mortality rate was 2.3 per 1000, compared to a mortality rate of 176.5 per 1000 when the last test was a nonreactive, positive CST. Reported false-positive rates for the CST range from 8% to 57%, with an average of approximately 30%. Principal advantages of this form of testing include excellent sensitivity and a weekly testing interval. Limitations include a high rate of equivocal results requiring repeat testing, increased expense and inconvenience (particularly if oxytocin is required), and increased time requirement compared to the NST. Additionally, use of the CST is contraindicated in several clinical settings, including preterm labor, placenta previa, vasa previa, cervical incompetence, multiple gestation, and previous classic cesarean section.
Biophysical Profile
The biophysical profile (BPP) assesses five biophysical variables: FHR reactivity, fetal movement, tone, and breathing reflect acute CNS function, while amniotic fluid volume serves as a marker of the longer-term adequacy of placental function. Two points are assigned for each normal variable and zero for each abnormal variable for a maximum score of 10. A BPP score of 8 to 10, with normal amniotic fluid volume, is considered normal. A score of 6 is considered suspicious, and testing usually is repeated the following day. Scores less than 6 are associated with increased perinatal morbidity and mortality; they usually warrant hospitalization for further evaluation or delivery. Among 12,620 women tested weekly using the BPP, Manning and co-workers reported a false-negative rate of 0.6 per 1000. The false-positive rate of the BPP varies with the score of the last test prior to delivery. Manning and co-workers reported a false-positive rate of zero among 11 patients in whom the last BPP score before delivery was zero, compared to a false-positive rate greater than 40% among 182 patients with a last BPP score of 6. The BPP is a reliable predictor of fetal well-being. The false-negative rate is superior to that of the NST alone and compares favorably with the false-negative rate of the CST. Advantages of BPP include excellent sensitivity, a weekly testing interval, and a low false-negative rate. The primary limitation is the requirement for personnel trained in sonographic visualization of the fetus. Additionally, although the duration of ultrasound observation is less than 10 minutes in the majority of cases, the complete BPP is more time-consuming than other noninvasive tests.
Modified Biophysical Profile
The modified biophysical profile (MBPP) combines the strengths of the NST (ease of use, low cost) and the complete BPP (improved sensitivity, low false-negative rate), while minimizing the requirement for additional training in sonographic visualization of the fetus. The test is performed once to twice weekly and uses the NST as a short-term marker of fetal status and the amniotic fluid index (AFI) as a marker of longer-term placental function. Interpretation of the NST incorporates assessment of reactivity, baseline rate, variability, and FHR decelerations. Late, prolonged, or significant variable decelerations, particularly in the setting of borderline amniotic fluid volume (AFI 5 to 10 cm), are considered abnormal. Regardless of reactivity, oligohydramnios (AFI < 5 cm) constitutes an abnormal test. Nageotte and co-workers (1994) evaluated 2774 high-risk pregnancies with twice weekly MBPPs and reported one unexplained fetal death within 1 week of a normal test result, for a false-negative rate of 0.36 per 1000. Miller and colleagues (1996) reported 54,617 MBPPs in 15,482 high-risk pregnancies. Antepartum testing in high-risk pregnancies yielded a fetal death rate that was nearly sevenfold lower than that in the untested, low-risk population. The overall false-negative rate of the MBPP was 0.8 per 1000, and the false-positive rate was 60%. Abnormal test results prompted intervention in 15.5% of the tested population; however, iatrogenic prematurity occurred in only 1.5% of women tested before 37 weeks. The false-negative rate of the MBPP is similar to the CST and the complete BPP. Additionally, it is easier to perform and less time-consuming than the CST or the complete BPP. The sensitivity of the MBPP is superior to that of the NST alone.
Limitations include the need for backup testing in 10% to 50% of patients, a high false-positive rate, and a twice weekly testing interval.
Doppler Velocimetry
Doppler velocimetry of fetal, umbilical, and uterine vessels has been the focus of intensive study in recent years. This technology uses systolic-to-diastolic flow ratios and resistance indices to estimate blood flow in various arteries. Studies have shown statistically significant improvement in perinatal outcome with the use of Doppler ultrasonography in pregnancies complicated by fetal growth restriction. Although severe restriction of umbilical artery blood flow, as evidenced by absent or reversed flow during diastole, has been correlated with fetal growth restriction, acidosis, and adverse perinatal outcome, the predictive values of less extreme deviations from normal remain undefined. In conditions other than fetal growth restriction, Doppler velocimetry does not appear to be a useful screening test for the detection of fetal compromise and is not recommended for use as a screening test in the general obstetric population. Doppler velocimetry is used in some settings as an adjunct to standard methods of fetal assessment but should not be considered a replacement for traditional fetal monitoring.
OTHER METHODS AND INTRAPARTUM EVALUATIONS
Fetal Scalp Blood Sampling
Fetal scalp blood sampling allows for the determination of the fetal acid–base status during labor. The technique requires dilation of the cervix, rupture of the membranes, and access to the fetal presenting part. A lighted plastic endoscopic cone is inserted into the vagina and through the cervical os so that it rests against the fetal presenting part. Care should be taken to ensure that it is not placed over a fontanelle. The area to be sampled is dried with a sponge and coated with a thin layer of silicone to facilitate the formation of a blood globule. The scalp is then punctured with a microscalpel, and blood is collected by capillary action in a heparinized capillary tube. After the sample is mixed, it is transported on ice to the laboratory for blood gas analysis. During labor, a normal scalp blood pH is 7.25 to 7.35. A fetal scalp pH of greater than or equal to 7.25 provides evidence of a nonacidotic fetus. A scalp pH of 7.20 to 7.25 is considered suspicious, and sampling should be repeated within 30 to 60 minutes. Historically, pH values less than 7.20 have been considered acidotic; however, minor deviations below normal correlate poorly with perinatal outcome. Because abnormal perinatal outcome has not been consistently observed with values greater than 7.0, there is debate regarding the specific pH value that should be considered acidotic. In addition, fetal scalp sampling reflects the status of the peripheral blood, where acidosis is inherent, owing to the accumulation of CO2. Since respiratory acidemia is generated in the blood and metabolic acidemia is generated in the tissues, a scalp sample may not reflect the state of the fetus. In light of the technical difficulty of the procedure and the uncertainty regarding interpretation of results, many centers have reduced their reliance on fetal scalp blood sampling.
Percutaneous Umbilical Blood Sampling
Electronic FHR monitoring, ultrasound, and fetal scalp blood sampling can provide useful information regarding the acid–base status of the fetus. Occasionally, however, direct access to circulating fetal blood is necessary. A classic example is the fetus with severe anemia secondary to Rh-isoimmunization. In this condition, maternal antibodies directed against fetal red blood cell antigens cross the placenta and bind to fetal red blood cells, resulting in hemolysis. At term, suspected fetal anemia can be managed by delivery and treatment of the neonate. Earlier in pregnancy, however, it may be preferable to delay delivery by assessing the fetal hematocrit and, if necessary, performing intrauterine blood transfusion. Percutaneous umbilical blood sampling (PUBS) is a procedure that affords direct access to fetal venous blood. Using sterile technique and direct ultrasound guidance, a fine needle is passed transabdominally into the umbilical vein. Medications or blood may be infused through the needle once fetal blood samples have been obtained. Other indications for PUBS include suspected antibody-mediated fetal thrombocytopenia and fetal cardiac arrhythmias requiring assessment of fetal drug levels or direct fetal administration of antiarrhythmic agents.
Fetal blood sampling can be used along with cardiotocography (CTG) to assess fetal acid–base status during labor. However it requires additional expertise, is time-consuming, has some significant risks, and has only intermittent information. For these reasons, it is not widely used.
Fetal Scalp Stimulation and Fetal Vibroacoustic Stimulation
FHR accelerations of 15 beats per minute for 15 seconds in response to fetal scalp stimulation have been shown to predict a scalp pH greater than or equal to 7.19. Among fetuses without an acceleratory response to scalp stimulation, 39% were considered acidotic (pH < 7.19). A similar relationship has been reported between fetal scalp pH and the FHR response to vibroacoustic stimulation with an artificial larynx applied to the maternal abdomen over the fetal head for 1 to 3 seconds. Among 30 fetuses with FHR accelerations in response to this stimulus, all had scalp pH values greater than or equal to 7.25. Half of the fetuses that did not respond to acoustic stimulation had pH values less than 7.25. FHR accelerations in response to external stimuli are thought to have the same predictive value as spontaneous accelerations. Fetal stimulation is used in antepartum testing to shorten the time of the NST and in the intrapartum period to confirm fetal well-being when spontaneous accelerations are absent. There is no evidence in humans of adverse long-term effects of vibroacoustic stimulation.
Fetal Pulse Oximetry (Fetal Oxygenation)
The aim of fetal pulse oximetry is to provide a continuous assessment of fetal oxygen saturation. The fetal pulse oximeter provides a noninvasive continuous measurement of fetal arterial oxyhemoglobin saturation once membranes have been ruptured. A sensor, similar to an IUPC is placed transvaginally between the uterine wall and fetal face and measures oxyhemoglobin saturation. The system includes an optoelectronic sensor and a microprocessor-based monitor. It has two low-voltage, light-emitting diodes as light sources and one photodetector. One of the light-emitting diodes emits red light (735 nm) and the other emits infrared light (890 nm). When light from each light-emitting diode passes through fetal tissues at the sensor application site, a fraction is absorbed. The photodetector measures the reflected (nonabsorbed) light using a process similar to measurement of transmitted light in conventional pulse oximetry. The amount of light absorbed is related to the amount of oxy- and deoxyhemoglobin. Their ratio determines fetal oxygen saturation during each arterial pulse. The potential is that this technology will provide a practical and easy method for better evaluation of fetal well-being during labor.
In 1994, fetal oximetry was marketed in countries outside of the U.S., including Europe and Canada. In January 2000, the Obstetrics and Gynecology Devices Panel of the Medical Devices Advisory Committee of the Food and Drug Administration (FDA) recommended approval of a fetal oximeter for use in obstetrics (OxiFirst: N-400 Fetal Oxygen Monitoring System, Nellcor).
A U.S. multicenter, randomized, controlled trial of fetal oximetry enrolled 1010 women with abnormal FHR patterns. The objective of the trial was to determine whether fetal pulse oximetry with fetal monitoring would reduce the rate of cesarean delivery performed for non-reassuring fetal status. The rate of cesarean delivery for fetal distress was significantly lower in the fetal oximetry group (4.5%) compared to 10.2% in the conventional fetal monitoring group. Unexpectedly, however, the cesarean rate for dystocia was significantly higher in the oximetry group (18.5%) versus 8.6% in the control group), thus the overall cesarean rates were not different (29% oximetry plus FHM vs. 26% FHM).
Clearly, continuous measurement of fetal oxygen saturation could be an important addition to the interpretation of heart rate patterns. The device is attractive in that it offers a direct measure of fetal oxygenation during labor. However, like its predecessor, electronic FHR monitoring, oximetry has the potential for widespread application without conclusive evidence that it is efficacious. The use of this technology is the focus of several ongoing clinical studies and trials.
Fetal Cardiotocography Plus Pulse Rate-Interval Analysis
Normally there is a negative relationship between the pulse rate (PR) interval and the FHR: as the FHR slows, the PR interval lengthens, and vice versa. In acidemic infants, this relationship is reversed. A retrospective study of 265 women suggested that the addition of time-interval analysis of the fetal ECG would decrease the rate of unnecessary fetal blood sampling or assisted delivery for presumed fetal distress. However, in a prospective randomized trial of the use of fetal ECG time-interval variables in addition to cardiotocography in fetal surveillance during labor there was no difference in operative intervention or neonatal outcome. The use of fetal cardiotocography plus PR-interval analysis is presently considered investigational.
ST Waveform Analysis
Since different fetuses may be able to tolerate varied oxygenation levels based on their situation prior to the hypoxic event, reliance on the level of oxygenation may not be the most applicable measure. In the search for other parameters, an evaluation of the changes in specific end organs, such as the fetal heart, have been pursued. One development is the ST waveform analysis to aid in the interpretation of ominous FHR patterns.
ST analysis of the ECG during exercise is a method for assessing myocardial function in the adult. As a corollary, ST waveform analysis of the fetus during labor is analogous to a fetal “stress test” and should provide information on the ability of the fetal heart to respond. An ST segment rise indicates a fetus responding to hypoxia. A negative ST indicates a fetus that is unable to respond or has not had time to react.
The fetal ECG is readily obtainable during labor from the same scalp electrode used to obtain the FHR. The evidence from experimental work indicates that ST waveform elevation reflects compensated myocardial stress and a switch to anaerobic myocardial metabolism. A study of 4966 women with CTG versus CTG plus ST analysis revealed that intrapartum monitoring with cardiotocography combined with automatic ST waveform analysis increased the ability of the obstetrician to identify fetal hypoxia. Currently, the use of ST segment elevation in clinical interpretation of fetal heart tracings is considered investigational and studies are ongoing to assess its effectiveness.
SUMMARY POINTS
· Fetal evaluation and interpretation of the findings has made some significant advances. However, there are many limitations with the available technology.
· Electronic FHR monitoring is a very sensitive tool for the detection of fetal compromise; truly compromised fetuses rarely fail to exhibit abnormal FHR patterns. The converse, however, is not true. Abnormal FHR patterns frequently are observed in the absence of fetal compromise. The limited positive predictive value is the principal shortcoming of FHR monitoring.
· Accuracy may be improved by combining FHR analysis with assessment of biophysical variables such as amniotic fluid volume, fetal movement, breathing, tone, and blood flow characteristics.
· Other variables under investigation include continuous intrapartum fetal pulse oximetry, fetal ECG interpretation, and ST waveform analysis.
· To date, the most effective combination of variables has not been defined, and no one approach to fetal surveillance has demonstrated clear superiority over the others. Yet, despite the limitations, antepartum testing in “high-risk” pregnancies has been reported to yield a fetal death rate nearly seven times lower than that in untested, “low-risk” pregnancies. If this observation is substantiated, future investigation will be needed to address the role of antepartum fetal surveillance in uncomplicated, low-risk pregnancies.
SUGGESTED READINGS
American College of Obstetricians and Gynecologists. Neonatal encephalopathy and cerebral palsy: defining the pathogenesis and pathophysiology, a report. Washington: American College of Obstetricians and Gynecologists, 2003.
Asphyxia and Cerebral Palsy
Nelson KB, Ellenberg JH. Antecedents of cerebral palsy: multivariate analysis of risk. N Engl J Med 1986;315:81.
Fetal Heart Rate Monitoring
American College of Obstetricians and Gynecologists. ACOG Technical Bulletin No. 207, July 1995.
Hon EH, Quilligan EJ. The classification of fetal heart rate. Conn Med 1967;31:779.
Kubli FW, Hon EH, Khazin AF, et al. Observations on heart rate and pH in the human fetus during labor. Am J Obstet Gynecol 1969;104:1190.
Paul RH, Suidan AK, Yeh S, et al. The evaluation and significance of intrapartum baseline FHR variability. Am J Obstet Gynecol 1975;123:206.
Rosen KG, Luzietti R. Intrapartum fetal monitoring: its basis and current developments. Prenatal Neonatal Med 2000;5:155–168.
Fetal Scalp Stimulation/Vibroacoustic Stimulation
Smith CV, Nguyen HN, Phelan JP, et al. Intrapartum assessment of fetal well-being: a comparison of fetal acoustic stimulation with acid-base determinations. Am J Obstet Gynecol 1986;155:726.
Prolonged Decelerations in Fetal Heart Rate Baseline
Clapp JF, Peress NS, Wesley M, et al. Brain damage after intermittent partial cord occlusion in the chronically instrumented fetal lamb. Am J Obstet Gynecol1988;159:504.
Contraction Stress Test/Oxytocin Challenge Test
Nonstress Test
Boehm FH, Salyer S, Shah DM, et al. Improved outcome of twice weekly nonstress testing. Obstet Gynecol 1986;67:566.
Freeman RK, Anderson G, Dorchester W. A prospective multi-institutional study of antepartum fetal heart rate monitoring. II. Contraction stress test versus nonstress test for primary surveillance. Am J Obstet Gynecol 1982;143:778.
Manning FA, Lange IR, Morrison I, Harman CR. Determination of fetal health: methods for antepartum and intrapartum fetal assessment. Curr Probl Obstet Gynecol 1983;7:3.
Yanagihara T, Ueta M, Hanaoka U, et al. Late second trimester nonstress test characteristics in preterm delivery before 32 weeks of gestation. Gynecol Obstet Invest 2001;1:32–35.
Modified Biophysical Profile
Clark SL, Sabey P, Jolley K. Nonstress testing with acoustic stimulation and amniotic fluid assessment: 5973 tests without unexpected fetal death. Am J Obstet Gynecol 1989;160:694.
Lagrew DC, Pircon RA, Nageotte M, et al How frequently should the amniotic fluid index be repeated? Am J Obstet Gynecol 1992;167:1129.
Miller DA, Rabello YA, Paul RH. The modified biophysical profile: ante-partum testing in the 1990's. Am J Obstet Gynecol 1996;174:812.
Nageotte JP, Towers CV, Asrat T, et al. Perinatal outcome with the MBPP. Am J Obstet Gynecol 1994;170:1672.
Fetal Scalp Blood Sampling
Goodwin TM, Milner-Masterson L, Paul RH. Elimination of fetal scalp blood sampling on a large clinical service. Obstet Gynecol 1994;83:971.
Doppler Ultrasound
Alfirevic Z, Neilson JP. Doppler ultrasonography in high-risk pregnancies: systematic review with meta-analysis. Am J Obstet Gynecol 1995;172:1379.
Fetal Pulse Oximetry
Garite TJ, Dildy GA, McNamara H, et al. A multi-center controlled trial of fetal pulse oximetry in the intrapartum management of nonreassuring fetal heart rate patterns. Am J Obstet Gynecol 2000;183: 1049–1058.
Fetal Cardiotocography Plus Pulse Rate-Interval Analysis
Strachan BK, van Wijngaarden WJ, Sahota D, et al. Cardiotocography only versus cardiotocography plus PR-interval analysis in intrapartum surveillance: a randomized, multicenter trial. Lancet 2000 355:456–459.
Strachan B, Sahota D, van Wijngaarden WJ, et al. The fetal electrocardiogram: relationship with academia at delivery. Am J Obstet Gynecol 2000;182:603–606.
ST Waveform Analysis
Amer-Wahlin I, Hellsten C, Noren H, et al. Cardiotocography only versus cardiotocography plus ST analysis of fetal electrocardiogram for intrapartum fetal monitoring: a Swedish randomized controlled trial. Lancet 2001 358: 534–538.