PART III
CHAPTER 14
Karen S. Clark , Teresa G. Hippel
OUTLINE
Manual Cell Counts
Equipment
Calculations
White Blood Cell Count
Platelet Count
Red Blood Cell Count
Disposable Blood Cell Count Dilution Systems
Body Fluid Cell Counts
Hemoglobin Determination
Principle
Microhematocrit
Rule of Three
Red Blood Cell Indices
Mean Cell Volume
Mean Cell Hemoglobin
Mean Cell Hemoglobin Concentration
Reticulocyte Count
Principle
Absolute Reticulocyte Count
Corrected Reticulocyte Count
Reticulocyte Production Index
Reticulocyte Control
Automated Reticulocyte Count
Erythrocyte Sedimentation Rate
Principle
Modified Westergren Erythrocyte Sedimentation Rate
Wintrobe Erythrocyte Sedimentation Rate
Disposable Kits
Automated Erythrocyte Sedimentation Rate
Point-of-Care Testing
Point-of-Care Tests
Objectives
After completion of this chapter, the reader will be able to:
CASE STUDIES
After studying the material in this chapter, the reader should be able to respond to the following case studies:
Case 1
The following results are obtained for a patient with normocytic, normochromic red blood cells on a peripheral blood film:
RBC count = 4.63 × 1012/L
HGB = 15 g/dL
HCT = 40% (0.40 L/L)
Case 2
For another patient, the following results are obtained:
RBC count = 3.20 × 1012/L
HGB = 5.8 g/dL
HCT = 18.9% (0.19 L/L)
Case 3
The following results are obtained for a patient using a point-of-care device that employs the conductivity method to measure the hematocrit:
Sodium = 160 mmol/L (Reference interval: 135 to 145 mmol/L)
Potassium = 3.6 mmol/L (Reference interval: 3.5 to 5.5 mmol/L)
HCT = 17.0% (0.17 L/L)
HGB = 6.0 g/dL
Clinical laboratory hematology has evolved from simple observation and description of blood and its components to a highly automated, extremely technical science, including examination at the molecular level. However, some of the more basic tests have not changed dramatically over the years. This chapter provides an overview of these basic tests and presents the manual and semiautomated methods that can be used in lieu of automated instrumentation. Included in this chapter is a discussion of point-of-care testing in hematology.
Manual cell counts
Although most routine cell-counting procedures in the hematology laboratory are automated, it may be necessary to use manual methods when counts exceed the linearity of an instrument, when an instrument is nonfunctional and there is no backup, in remote laboratories in Third World countries, or in a disaster situation when testing is done in the field. Although the discussion in this chapter concerns whole blood, body fluid cell counts are also often performed using manual methods. Chapter 18 discusses the specific diluents and dilutions used for body fluid cell counts. Chapter 15 discusses automated cell-counting instrumentation in detail.
Manual cell counts are performed using a hemacytometer, or counting chamber, and manual dilutions made with calibrated, automated pipettes and diluents (commercially available or laboratory prepared). The principle for the performance of cell counts is essentially the same for white blood cells (WBCs), red blood cells (RBCs), and platelets; only the dilution, diluting fluid, and area counted vary. Any particle (e.g., sperm) can be counted using this system.
Equipment
Hemacytometer
The manual cell count uses a hemacytometer, or counting chamber. The most common one is the Levy chamber with improved Neubauer ruling. It is composed of two raised surfaces, each with a 3 mm × 3 mm square counting area or grid (total area 9 mm2), separated by an H-shaped moat. As shown in Figure 14-1, this grid is made up of nine 1 mm × 1 mm squares. Each of the four corner (WBC) squares is subdivided further into 16 squares, and the center square subdivided into 25 smaller squares. Each of these smallest squares is 0.2 mm × 0.2 mm which is 1/25 of the center square or 0.04 mm2. A coverslip is placed on top of the counting surfaces. The distance between each counting surface and the coverslip is 0.1 mm; thus the totalvolume of one entire grid or counting area on one side of the hemacytometer is 0.9 mm3. Hemacytometers and coverslips must meet the specifications of the National Bureau of Standards, as indicated by the initials “NBS” on the chamber. When the dimensions of the hemacytometer are thoroughly understood, the area counted can be changed to facilitate the counting of samples with extremely low or high counts.
FIGURE 14-1 Hemacytometer and a close-up view of the counting areas as seen under the microscope. The areas for the standard white blood cell count are labeled W, and the areas for the standard red blood cell count are labeled R. The entire center square, outlined in blue, is used for counting platelets. The side view of the hemacytometer shows a depth of 0.1 mm from the surface of the counting grid to the coverslip.
Calculations
The general formula for manual cell counts is as follows and can be used to calculate any type of cell count:
Or
*Reciprocal of depth.
The calculation yields the number of cells per mm3. One mm3 is equivalent to one microliter (μL). The count per μL is converted to the count per liter (L) by multiplying by a factor of 106.
White blood cell count
The WBC or leukocyte count is the number of WBCs in 1 liter (L) or 1 microliter (μL) of blood. Whole blood anticoagulated with ethylenediaminetetraacetic acid (EDTA) or blood from a skin puncture is diluted with 1% buffered ammonium oxalate or a weak acid solution (3% acetic acid or 1% hydrochloric acid). The diluting fluid lyses the nonnucleated red blood cells in the sample to prevent their interference in the count. The typical dilution of blood for the WBC count is 1:20. A hemacytometer is charged (filled) with the well-mixed dilution and placed under a microscope and the number of cells in the 4 large corner squares (4 mm2) is counted.
Procedure
FIGURE 14-2 One large corner square of a hemacytometer indicating which cells to count. Cells touching the left and top lines (solid circles) are counted. Cells touching bottom and right (open circles) are not counted.
FIGURE 14-3 White blood cells as seen in the hemacytometer under low power (10× objective) 100× total magnification.
BOX 14-1
How to Make a Moist Chamber
A moist chamber may be made by placing a piece of damp filter paper in the bottom of a Petri dish. An applicator stick broken in half can serve as a support for the hemacytometer.
Example using the first equation
When a 1:20 dilution is used, the four large squares on one side of the chamber yield counts of 23, 26, 22, and 21. The total count is 92. The four large squares on the other side of the chamber yield counts of 28, 24, 22, and 26. The total count is 100. The difference between sides is less than 10%.
The average number of cells of the two sides of the chamber is 96. Using the average in the formula:
Alternately, a 1:100 dilution may be used counting the number of cells in the entire counting area (9 large squares, 9 mm2) on both sides of the chamber (Table 14-1). As an example, if an average of 54 cells were counted in the entire counting area on both sides of the chamber:
TABLE 14-1
Manual Cell Counts with Most Common Dilutions, Counting Areas
General reference intervals for males and females in different age groups can be found on the inside front cover of this text. Reference intervals may vary slightly according to the population tested and should be established for each laboratory.
Sources of error and comments
Report the result as the “corrected” WBC count.
Platelet count
A platelet count is the number of platelets in 1 liter (L) or 1 microliter (μL) of whole blood. Platelets adhere to foreign objects and to each other, which makes them difficult to count. They also are small and can be confused easily with dirt or debris. In this procedure, whole blood, with EDTA as the anticoagulant, is diluted 1:100 with 1% ammonium oxalate to lyse the nonnucleated red blood cells. The platelets are counted in the 25 small squares in the large center square (1 mm2) of the hemacytometer using a phase-contrast microscope in the reference method described by Brecher and Cronkite.1 A light microscope can also be used, but visualizing the platelets may be more difficult.
Procedure
General reference intervals for males and females according to age groups can be found on the inside front cover of this text.
Sources of error and comments
Red blood cell count
Manual RBC counts are rarely performed because of the inaccuracy of the count and questionable necessity. Use of other, more accurate manual RBC procedures, such as the microhematocrit and hemoglobin concentration, is desirable when automation is not available.
Table 14-1 contains information on performing manual WBC, platelet, and RBC counts.
Disposable blood cell count dilution systems
Capillary pipette and diluent reservoir systems are commercially available for WBC and platelet counts. One such system is LeukoChek™ (Biomedical Polymers, Inc., Gardner, MA). It consists of a capillary pipette (calibrated to accept 20 μL of blood) that fits into a plastic reservoir containing 1.98 mL of 1% buffered ammonium oxalate (Figure 14-4). Blood from a well-mixed EDTA-anticoagulated specimen or from a skin puncture is allowed to enter the pipette by capillary action to the fill volume. The blood is added to the reservoir making a 1:100 dilution. After mixing the reservoir and allowing 10 minutes for lysis of the red blood cells, the reverse end of the capillary pipette is placed in the reservoir cap making a dropper. The first 3 or 4 drops of the diluted sample is discarded, and the capillary pipette is used to charge the hemacytometer.
FIGURE 14-4 LeukoChek™ blood diluting system for manual white blood cell and platelet counts. It consists of a 20 µL capillary pipette and plastic reservoir containing 1.98 mL of 1% buffered ammonium oxalate that makes a 1:100 dilution of whole blood. Source: (Courtesy Biomedical Polymers, Inc., Gardner, MA.)
Both WBC and platelet counts can be done from the same diluted sample. WBCs are counted in all 9 large squares (9 mm2) using low power (100× total magnification). Platelets are counted in the 25 small squares in the center square (1 mm2) using high power (400× total magnification). The standard formula is used to calculate the cell counts.
Body fluid cell counts
Body fluid cell counts are discussed in detail in Chapter 18.
Hemoglobin determination
The primary function of hemoglobin within the red blood cell is to carry oxygen to and carbon dioxide from the tissues. The cyanmethemoglobin (hemoglobincyanide) method for hemoglobin determination is the reference method approved by the Clinical and Laboratory Standards Institute.2
Principle
In the cyanmethemoglobin method, blood is diluted in an alkaline Drabkin solution of potassium ferricyanide, potassium cyanide, sodium bicarbonate, and a surfactant. The hemoglobin is oxidized to methemoglobin (Fe3+) by the potassium ferricyanide, K3Fe(CN)6. The potassium cyanide (KCN) then converts the methemoglobin to cyanmethemoglobin:
The absorbance of the cyanmethemoglobin at 540 nm is directly proportional to the hemoglobin concentration. Sulfhemoglobin is not converted to cyanmethemoglobin; it cannot be measured by this method. Sulfhemoglobin fractions of more than 0.05 g/dL are seldom encountered in clinical practice, however.3
Procedure
FIGURE 14-5 Standard curve obtained when a cyanmethemoglobin standard of 80 mg/dL is used. A blank (100% transmittance) and four dilutions were made: 5 g/dL (72.9% transmittance), 10 g/dL (53.2% transmittance), 15 g/dL (39.1% transmittance), and 20 g/dL (28.7% transmittance).
General reference intervals can be found on the inside cover of this text.
Sources of error and comments
FIGURE 14-19 The HemoCue ® Hb 201+ System for measuring hemoglobin. Source: (Courtesy HemoCue, Inc., Brea, CA.)
Microhematocrit
The hematocrit is the volume of packed red blood cells that occupies a given volume of whole blood. This is often referred to as the packed cell volume (PCV). It is reported either as a percentage (e.g., 36%) or in liters per liter (0.36 L/L).
Procedure
FIGURE 14-6 Microhematocrit reader.
FIGURE 14-7 Capillary tube with anticoagulated whole blood after it has been centrifuged. Notice the layers containing plasma, the buffy coat (white blood cells and platelets), and the red blood cells.
BOX 14-2
Determining Maximum Packing Time for Microhematocrit
The time to obtain maximum packing of red blood cells should be determined for each centrifuge. Duplicate microhematocrit determinations should be made using fresh, well-mixed blood anticoagulated with ethylenediaminetetraacetic acid (EDTA). Two specimens should be used, with one of the specimens having a known hematocrit of 50% or higher. Starting at 2 minutes, centrifuge duplicates at 30-second intervals and record results. When the hematocrit has remained at the same value for two consecutive readings, optimum packing has been achieved, and the second time interval should be used for microhematocrit determinations.10
Sources of error and comments
The READACRIT centrifuge (Becton, Dickinson and Company, Franklin Lakes, NJ) uses precalibrated capillary tubes and has built-in hematocrit scales, which eliminates the need for separate reading devices (Figure 14-8). The use of SUREPREP Capillary Tubes (Becton, Dickinson) eliminates the use of sealants. They have a factory-inserted plug that seals automatically when the blood touches the plug.11
FIGURE 14-8 READACRIT centrifuge with built-in capillary tube compartments and hematocrit scales. Source: (Courtesy and © Becton, Dickinson and Company, Franklin Lakes, NJ.)
Rule of three
When samples are analyzed by automated or manual methods, a quick visual check of the results of the hemoglobin and hematocrit can be done by applying the “rule of three.” This rule applies only to samples that have normocytic normochromic red blood cells. The value of the hematocrit should be three times the value of the hemoglobin plus or minus 3: HGB × 3 = HCT ± 3 (0.03 L/L). It should become habit for the analyst to multiply the hemoglobin by 3 mentally for every sample; a value discrepant with this rule may indicate abnormal red blood cells, or it may be the first indication of error.
For example, the following results are obtained from patients:
Case 1
HGB = 12 g/dL
HCT = 36% (0.36 L/L)
According to the rule of three,
An acceptable range for the hematocrit would be 33% to 39%. These values conform to the rule of three.
Case 2
HGB = 9 g/dL
HCT = 32%
According to the rule of three,
An acceptable range for hematocrit would be 24% to 30%, so these values do not conform to the rule of three.
Case 3
HGB = 15 g/dL
HCT = 36%
According to the rule of three,
An acceptable range for hematocrit would be 42% to 48%, so these values do not conform to the rule of three.
If values do not agree, the blood film should be examined for abnormal red blood cells; causes of false increases and decreases in the hemoglobin and/or hematocrit values should also be investigated. In the second example, the blood film reveals red blood cells that are low in hemoglobin concentration (hypochromic) and are smaller in volume (microcytic), so the rule of three cannot be applied. If red blood cells do appear normal, possible causes of a falsely low hemoglobin concentration or a falsely elevated hematocrit should be investigated. In the third example, the specimen is determined to have lipemic plasma causing a falsely elevated hemoglobin concentration, and a correction must be made to obtain an accurate hemoglobin value. (See Hemoglobin Determination in this chapter.)
When an unexplained discrepancy is found, the sample processed before and after the sample in question should be checked to determine whether they conform to the rule. If they do not conform, further investigation should be done to find the problem. A control sample should be run when such a discrepancy is found. If the instrument produces appropriate results for the control, random error may have occurred (Chapter 5).
Red blood cell indices
The mean cell volume (MCV), mean cell hemoglobin (MCH), and mean cell hemoglobin concentration (MCHC) are the RBC indices. These are calculated to determine the average volume and hemoglobin content and concentration of the red blood cells in the sample. In addition to serving as a quality control check, the indices may be used for initial classification of anemias. Table 14-2 provides a summary of the RBC indices, morphology, and correlation with various anemias. The morphologic classification of anemia on the basis of MCV is discussed in detail in Chapter 19.
TABLE 14-2
Red Blood Cell Indices, Red Blood Cell Morphology, and Disease States
Hb, Hemoglobin; MCHC, mean cell hemoglobin concentration; MCV, mean cell volume.
Mean cell volume
The MCV is the average volume of the red blood cell, expressed in femtoliters (fL), or 10−15 L:
The reference interval for MCV is 80 to 100 fL. RBCs with an MCV of less than 80 fL are microcytic; those with an MCV of more than 100 fL are macrocytic.
Mean cell hemoglobin
The MCH is the average weight of hemoglobin in a red blood cell, expressed in picograms (pg), or 10−12 g:
For example, if the hemoglobin = 16 g/dL and the RBC count = 5 × 1012/L, the MCH = 32 pg.
The reference interval for adults is 26 to 32 pg. The MCH generally is not considered in the classification of anemias.
Mean cell hemoglobin concentration
The MCHC is the average concentration of hemoglobin in each individual red blood cell. The units used are grams per deciliter (formerly given as a percentage):
For example, if the HGB = 16 g/dL and the HCT = 48%, the MCHC = 33.3 g/dL.
Values of normochromic red blood cells range from 32 to 36 g/dL; values of hypochromic cells are less than 32 g/dL, and values of “hyperchromic” cells are greater than 36 g/dL. Hypochromic red blood cells occur in thalassemias, iron deficiency, and other conditions listed in Table 14-2. The term hyperchromic is a misnomer: a cell does not really contain more than 36 g/dL of hemoglobin, but its shape may have become spherocytic, which makes the cell appear full. An MCHC between 36 and 38 g/dL should be checked for spherocytes. An MCHCgreater than 38 g/dL should be investigated for an error in hemoglobin value (see Sources of Error and Comments in the section on hemoglobin determination). Another cause for a markedly increased MCHC could be the presence of a cold agglutinin. Incubating the specimen at 37° C for 15 minutes before analysis usually produces accurate results. Cold agglutinin disease is discussed in more detail in Chapter 26.
Reticulocyte count
The reticulocyte is the last immature red blood cell stage. Normally, a reticulocyte spends 2 days in the bone marrow and 1 day in the peripheral blood before developing into a mature red blood cell. The reticulocyte contains remnant cytoplasmic ribonucleic acid (RNA) and organelles such as the mitochondria and ribosomes (Chapter 8). The reticulocyte count is used to assess the erythropoietic activity of the bone marrow.
Principle
Whole blood, anticoagulated with EDTA, is stained with a supravital stain, such as new methylene blue. Any nonnucleated red blood cell that contains two or more particles of blue-stained granulofilamentous material after new methylene blue staining is defined as a reticulocyte (Figure 14-9).
FIGURE 14-9 Reticulocytes with new methylene blue vital stain (peripheral blood ×1000). Reticulocytes are nonnucleated red blood cells with two or more blue-stained filaments or particles.
Procedure
For example, if 15 reticulocytes are counted,
Or the number of reticulocytes counted can be multiplied by 0.1 (100/1000) to obtain the result.
Miller disc
Because large numbers of red blood cells should be counted to obtain a more precise reticulocyte count, the Miller disc was designed to reduce this labor-intensive process. The disc is composed of two squares, with the area of the smaller square measuring 1/9 the area of the larger square. The disc is inserted into the eyepiece of the microscope and the grid in Figure 14-10 is seen. RBCs are counted in the smaller square, and reticulocytes are counted in the larger square. Selection of the counting area is the same as described earlier. A minimum of 112 cells should be counted in the small square, because this is equivalent to 1008 red cells in the large square and satisfies the College of American Pathologists (CAP) hematology standard for a manual reticulocyte count based on at least 1000 red cells.13 The calculation formula for percent reticulocytes is
FIGURE 14-10 Miller ocular disc counting grid as viewed through a microscope. The area of square B is 1/9 the area of square A. Alternatively, square B may be in the center of square A.
For example, if 15 reticulocytes are counted in the large square and 112 red blood cells are counted in the small square,
Equation reference interval
General reference intervals can be found on the inside front cover of this text.
Sources of error and comments
FIGURE 14-11 Reticulocytes (A) and Heinz bodies (B) stained with supravital stain (peripheral blood ×1000).
Absolute reticulocyte count
Principle
The absolute reticulocyte count (ARC) is the actual number of reticulocytes in 1 liter (L) or 1 microliter (μL) of blood.
Calculations
For example, if a patient’s reticulocyte count is 2% and the RBC count is 2.20 × 1012/L, the ARC is calculated as follows (note that the calculated result has to be converted from 1012/L to 109/L):
The absolute reticulocyte count can also be reported as the number of cells per μL. Using the example above, the RBC count in μL (2.20 × 106/μL) is used in the formula, and the ARC result is 44 × 103/μL.
Reference interval
Values between 20 × 109/L and 115 × 109/L are within the reference interval for most populations.14
Corrected reticulocyte count
Principle
In specimens with a low hematocrit, the percentage of reticulocytes may be falsely elevated because the whole blood contains fewer red blood cells. A correction factor is used, with the average normal hematocrit considered to be 45%.
Calculation
Reference interval
Patients with a hematocrit of 35% should have an elevated corrected reticulocyte count of 2% to 3% to compensate for the mild anemia. In patients with a hematocrit of less than 25%, the count should increase to 3% to 5% to compensate for the moderate anemia. The corrected reticulocyte count depends on the degree of anemia.
Reticulocyte production index
Principle
Reticulocytes that are released from the marrow prematurely are called shift reticulocytes. These reticulocytes are “shifted” from the bone marrow to the peripheral blood earlier than usual to compensate for anemia. Instead of losing their reticulum in 1 day, as do most normal circulating reticulocytes, these cells take 2 to 3 days to lose their reticula. When erythropoiesis is evaluated, a correction should be made for the presence of shift reticulocytes if polychromasia is reported in the red blood cell morphology. Most normal (nonshift) reticulocytes become mature red blood cells within 1 day after entering the bloodstream and thus represent 1 day’s production of red blood cells in the bone marrow. Cells shifted to the peripheral blood prematurely stay longer as reticulocytes and contribute to the reticulocyte count for more than 1 day. For this reason, the reticulocyte count is falsely increased when polychromasia is present, because the count no longer represents the cells maturing in just 1 day. On many automated instruments, this mathematical adjustment of the reticulocyte count has been replaced by the measurement of immature reticulocyte fraction (Chapter 15).12
The patient’s hematocrit is used to determine the appropriate correction factor (reticulocyte maturation time in days):
|
Patient’s Hematocrit Value (%) |
Correction Factor (Maturation Time, Days) |
|
40–45 |
1 |
|
35–39 |
1.5 |
|
25–34 |
2 |
|
15–24 |
2.5 |
|
< 15 |
3 |
Calculation
The reticulocyte production index (RPI) is calculated as follows:
Or
For example, for a patient with a reticulocyte count of 7.8% and a HCT of 30%, and with polychromasia noted, the previous table indicates a maturation time of 2 days. Thus
Reference interval
An adequate bone marrow response usually is indicated by an RPI that is greater than 3. An inadequate erythropoietic response is seen when the RPI is less than 2.14
Reticulocyte control
Several commercial controls are now available for monitoring manual and automated reticulocyte counts [e.g., Retic-Chex II, Streck Laboratories, Omaha, NE; Liquichek Reticulocyte Control (A), Bio-Rad Laboratories, Hercules, CA]. Most of the controls are available at three levels. The control samples are treated in the same manner as the patient samples. The control can be used to verify the laboratorian’s accuracy and precision when manual counts are performed.
Automated reticulocyte counts
The major instrument manufacturers offer are analyzers that perform automated reticulocyte counts. All of the analyzers evaluate reticulocytes using optical scatter or fluorescence after the red blood cells are treated with fluorescent dyes or nucleic acid stains to stain residual RNA in the reticulocytes. The percentage and the absolute count are provided. These results are statistically more valid because of the large number of cells counted. Other reticulocyte parameters that are offered on some automated instruments include a maturation index/immature reticulocyte fraction or IRF (reflecting the proportion of the more immature reticulocytes in the sample), the reticulocyte hemoglobin concentration, and reticulocyte indices (such as the mean reticulocyte volume and distribution width). The IRF may be especially useful in detecting early erythropoietic activity after chemotherapy or hematopoietic stem cell transplantation. The reticulocyte hemoglobin is useful to detect early iron deficiency (Chapter 20). Automated reticulocyte counting is discussed in Chapter 15.
Erythrocyte sedimentation rate
The erythrocyte sedimentation rate (ESR) is ordered with other tests to detect and monitor the course of inflammatory conditions such as, rheumatoid arthritis, infections, or certain malignancies. It is also useful in the diagnosis of temporal arteritis and polymyalgia rheumatica.15 The ESR, however, is not a specific test for inflammatory diseases and is elevated in many other conditions such as plasma cell myeloma, pregnancy, anemia, and older age. It is also prone to technical errors that can falsely elevate or decrease the sedimentation rate. Because of its low specificity and sensitivity, the ESR is not recommended as a screening test to detect inflammatory conditions in asymptomatic individuals.15 Other tests for inflammation, such as the C-reactive protein level, may be a more predictable and reliable alternative to monitor inflammation.16
Principle
When anticoagulated blood is allowed to stand at room temperature undisturbed for a period of time, the red blood cells settle toward the bottom of the tube. The ESR is the distance in millimeters that the red blood cells fall in 1 hour. The ESR is affected by red blood cell, plasma, and mechanical and technical factors. Red blood cells have a net negative surface charge and tend to repel one another. The repulsive forces are partially or totally counteracted if there are increased quantities of positively charged plasma proteins. Under these conditions the red blood cells settle more rapidly as a result of the formation of rouleaux (stacking of red blood cells). Examples of macromolecules that can produce this reaction are fibrinogen, β-globulins, and pathologic immunoglobulins.17, 18
Normal red blood cells have a relatively small mass and settle slowly. Certain diseases can cause rouleaux formation, in which the plasma fibrinogen and globulins are altered. This alteration changes the red blood cell surface, which leads to stacking of the red blood cells, increased red blood cell mass, and a more rapid ESR. The ESR is directly proportional to the red blood cell mass and inversely proportional to plasma viscosity. Several methods, both manual and automated, are available for measuring the ESR. Only the most commonly used methods are discussed here.
Modified westergren erythrocyte sedimentation rate
The most commonly used method today is the modified Westergren method. One advantage of this method is that the taller column height allows the detection of highly elevated ESRs. It is the method recommended by the International Council for Standardization in Hematology and the Clinical and Laboratory Standards Institute.15, 19
Procedure
FIGURE 14-12 Erythrocyte sedimentation rate (ESR), 1 hour = 93 mm, which is elevated above the reference intervals.
Wintrobe erythrocyte sedimentation rate
When the Wintrobe method was first introduced, the specimen used was oxalate-anticoagulated whole blood. This was placed in a 100-mm column. Today, EDTA-treated or citrated whole blood is used with the shorter column. The shorter column height allows a somewhat increased sensitivity in detecting mildly elevated ESRs.
Procedure
Reference interval
Reference intervals according to sex and age can be found on the inside front cover of this text. Table 14-3 lists some of the factors that influence the ESR.
TABLE 14-3
Factors Affecting the Erythrocyte Sedimentation Rate (ESR)
|
Category |
Increased ESR |
Decreased ESR |
|
Blood proteins and lipids |
Hypercholesterolemia Hyperfibrinogenemia |
Hyperalbuminemia Hyperglycemia |
|
Red blood cells |
Anemia Macrocytosis |
Acanthocytosis Anisocytosis (marked) |
|
White blood cells |
Leukemia |
Leukocytosis (marked) |
|
Drugs |
Dextran Heparin |
Adrenocorticotropic hormone (corticotropin) Cortisone |
|
Clinical conditions |
Acute heavy metal poisoning Acute bacterial infections |
Cachexia Congestive heart failure |
|
Specimen handling |
Refrigerated sample not returned to room temperature |
Clotted blood sample Delay in testing |
|
Technique |
High room temperature Tilted ESR tube |
Bubbles in ESR column Low room temperature |
From American Society for Clinical Pathology/American Proficiency Institute: 2006 2nd Test Event—Educational Commentary—The Erythrocyte Sedimentation Rate and Its Clinical Utility. API is the proficiency testing group that provides testing materials to the American Society for Clinical Pathology. The educational commentary itself is written by ASCP. This reference can also be accessed at: http://www.api-pt.com/Reference/Commentary/2006Bcoag.pdf. Accessed November 5, 2014.
Sources of error and comments
Disposable kits
Disposable commercial kits are available for ESR testing (Figure 14-13). Several kits include safety caps for the columns that allow the blood to fill precisely to the zero mark. This safety cap makes the column a closed system and eliminates the error involved in manually setting the blood to the zero mark.
FIGURE 14-13 Sediplast (Polymedco) disposable sedimentation rate system. Source: (Courtesy Polymedco, Cortlandt Manor, NY.)
Automated erythrocyte sedimentation rate
There are several automated ESR systems available using the traditional Westergren and Wintrobe methods, as well as alternate methods such as centrifugation. The Ves-Matic system (Diesse, Inc., Hialeah, FL) is a bench-top analyzer designed to determine ESR by use of an optoelectronic sensor, which measures the change in opacity of a column of blood as sedimentation of blood progresses. Blood is collected in special Ves-Tec or Vacu-Tec tubes, which contain sodium citrate and are compatible with the Vacutainer system. These tubes are used directly in the instrument (Figure 14-14). Acceleration of sedimentation is achieved by positioning the tubes at an 18-degree angle in relation to the vertical axis. Results comparable with Westergren 1-hour values are obtained in 20 minutes.20
FIGURE 14-14 Two models of the Ves-Matic instruments for sedimentation rates: The Ves-Matic Easy (A) for up to 10 specimens (requiring special tubes) and the Ves-Matic Cube 30 (B) for up to 30 specimens, determining the ESR directly from EDTA tubes. Products with up to 190-specimen capacity are also available. Source: (Courtesy Diesse Inc., Hialeah, FL.)
Another automated ESR analyzer is the Sedimat 15 (Polymedco, Cortlandt Manor, NY), which uses the principle of infrared measurement. It is capable of testing one to eight samples randomly or simultaneously and provides results in 15 minutes (Figure 14-15).
FIGURE 14-15 Sedimat 15 (Polymedco) automated sedimentation rate system. Source: (Courtesy Polymedco, Cortlandt Manor, NY.)
The ESR STAT PLUS system (HemaTechnologies, Lebanon, NJ) is based on centrifugation. The advantages of this method are a smaller required sample volume and shorter testing time, which makes it more suitable for a pediatric patient population. The disadvantage of this method is the number of exacting preanalytical steps that must be strictly followed to prevent erroneous results. Compliance with these steps may be difficult to achieve consistently in a busy hematology laboratory.21
Additional methods
Additional manual and semi-automated methods are included in other chapters that are relevant to their clinical application. Examples include: Chapter 24 for the osmotic fragility test and qualitative and quantitative assays for glucose-6-phosphate dehydrogenase and pyruvate kinase activity; Chapter 27 for the solubility test for Hb S, hemoglobin electrophoresis (alkaline and acid pH), and unstable hemoglobin test; and Chapter 28 for the vital stain for hemoglobin H and the Kleihauer-Betke acid elution test for Hb F distribution in the RBCs.
Point-of-care testing
Point-of-care testing offers the ability to produce rapid and accurate results that help facilitate faster treatment, which can decrease patient length of stay. This testing is rarely performed by trained laboratory personnel; most often, it is carried out by nurses. Manufacturers have created analyzers with nonlaboratory operators in mind, but results obtained using these systems are still affected by preanalytic and analytical variables. The laboratory’s partnership with nursing is the key to success in any hospital’s point-of-care program.
Point-of-care testing is defined as diagnostic testing at or near the site of patient care. The Clinical Laboratory Improvement Amendments of 1988 (CLIA) introduced the concept of “testing site neutrality,” which means that regardless of where the diagnostic testing is performed or who performs the test, all testing sites must follow the same regulatory requirements based on the “complexity” of the test. Under CLIA, point-of-care testing (including physician-performed microscopy) is classified as “waived” or “moderately complex.” Tests are classified as waived if they are determined to be “simple tests with an insignificant risk of an erroneous result.” Point-of-care testing is commonly performed in hospital inpatient units, outpatient clinics, surgery centers, emergency departments, long-term care facilities, and dialysis units. For waived point-of-care testing, facilities are required to obtain a certificate of waiver, pay the appropriate fees, and follow the manufacturers’ testing instructions.22 For any point-of-care program to be successful, certain key elements must be present. Clear administrative responsibility, well-written procedures, a training program, quality control, proficiency testing, and equipment maintenance are essential for success. The first step is appointing a laboratory point-of-care testing coordinator. This person not only is the “go-to” person but is also an important liaison between the laboratory and nursing staff. The second step to ensuring a successful program is to create a multidisciplinary team with authority to impact all aspects of the POC program. This committee would have the authority to oversee the integrity and quality of the existing POC program and institute changes or new testing as needed. It is also important to have administrative support to help remove barriers.
A point-of-care testing program must incorporate all of the following. A written policy should be developed that defines the program. This policy should outline who is responsible for each part of the program. The policy should also indicate where the testing is to be performed and who is going to perform the testing. Testing procedures should be written that clearly state how to perform the tests and that address how to handle critical values and/or any discrepant results. The program must be monitored. An ongoing evaluation of the point-of-care testing is vital for success.
When the instrument to be used in the point-of-care testing program is being selected, it is helpful to invite the vendors to demonstrate their equipment. An equipment display that is available for hands-on use by the operators can be very helpful in selection of the appropriate instrumentation. Patient correlation studies are very useful in choosing equipment that best covers the patient population for that particular institution. Point-of-care operators need handheld analyzers that are lightweight, accurate, fast, and that require little specimen material. The point-of-care testing system should also address the following laboratory concerns:
Paramount to point-of-care testing is patient safety. It is important to maintain good practices, and with waived testing, this often comes down to the basics. Such basics include proper and appropriate specimen collection, proper identification of the patient and specimen, proper storage of reagents, and good documentation of patient test results (use of point-of-care interfaces is beneficial), as well as proper performance of any necessary instrument maintenance. Laboratory oversight is sometimes absent, and basic safety precautions necessary for waived tests can be easily overlooked, often due to a lack of understanding, lack of training, and high personnel turnover rates.23 Patient safety, risk management, and error reduction are primary goals of all health care facilities. All testing personnel should be properly trained in best practices to avoid exposure. The individual responsible for oversight—whether laboratory or nonlaboratory—must avoid taking safety for granted. All applicable standards (including those of the Occupation Safety and Health Administration, Centers for Disease Control and Prevention, The Joint Commission, CAP, CLIA, and so forth) should be implemented and easily accessible. Because the number of waived tests has grown significantly since waived tests were first defined by CLIA, it is paramount that standard safety precautions and the basic steps outlined earlier be implemented to ensure that patient safety is not sacrificed in the unique situation of CLIA-waived testing.
Point-of-care tests
Various point-of-care instruments are available to measure parameters such as hemoglobin level and hematocrit, and some perform a complete blood count.
Hematocrit
The most common methods for determining the hematocrit include the microhematocrit centrifuge, conductometric methods, and calculation by automated cell counters (Chapter 15).
Centrifuge-based microhematocrit systems have been available for years, and the results obtained correlate well with the results produced by standard cell counters. Nonlaboratorians and inexperienced operators, however, may be unaware of the error that can be introduced by insufficient centrifugation time and inaccurate reading of the microhematocrit tube (see comments in the Microhematocrit section). Examples of centrifuge-based devices are the Hematastat II (Separation Technology, Inc., Altamonte Springs, FL) and STAT Crit (Wampole Laboratories, Cranbury, NJ).
The i-STAT 1 (Abbott Laboratories, Abbott Park, IL)24 (Figure 14-16) and the Epoc (Epocal, Inc., Ottawa, ON) (Figure 14-17)25 use the conductivity method to determine the hematocrit. Plasma conducts electrical current, whereas WBCs act as insulators. In the i-STAT system, before the measured sample conductance is converted into the hematocrit value, corrections are applied for the temperature of the sample, the size of the fluid segment being measured, and the relative conductivity of the plasma component. The first two corrections are determined from the measured value of the calibrant conductance and the last correction from the measured concentrations of sodium and potassium in the sample.24
FIGURE 14-16 i-STAT instrument for measuring hematocrit. Source: (Courtesy Abbott Laboratories, Abbott Park, IL.)
FIGURE 14-17 Epoc device for measuring hematocrit. Source: (Courtesy Epocal, Inc., Ottawa, Ontario, Canada.)
Sources of error and comments.
Conductivity of a whole blood sample is dependent on the amount of electrolytes in the plasma portion. Conductivity does not distinguish red blood cells from other nonconductive elements such as proteins, lipids, and WBCs that may be present in the sample.
A low total protein level will falsely decrease the hematocrit. The presence of lipids can interfere with the hematocrit measurement. An increased WBC count will falsely increase the hematocrit. The presence of cold agglutinins can falsely decrease the hematocrit.24
Other instruments.
Other instruments that measure the hematocrit include the following:
FIGURE 14-18 Gem Premier instrument for measuring hematocrit. Source: (Courtesy Instrumentation Laboratory Company, Lexington, MA.)
Hemoglobin concentration
In point-of-care testing, hemoglobin concentration is measured by modified hemoglobinometers or by oximeters integrated with a blood gas analyzer. The HemoCue hemoglobinometer (HemoCue, Inc., Brea, CA) uses a small cuvette that contains a lysing agent and reagents to form a hemoglobin azide, which is measured by a photometer at two wavelengths (570 nm and 880 nm) (Figure 14-19).6 This eliminates interference from turbidity in the sample. Results obtained with the instrument compare well with those produced by reference methods, but a major source of error is mixture of blood with tissue fluid during skin puncture collection. The AVOX 1000E (ITC) measures total hemoglobin by a spectrophotometric method. The STAT-Site MHbg Meter (Stanbio Laboratory, Boerne, TX) uses the azidemethemoglobin principle and reflectance photometry to measure reflected light in the test area. The test card is composed of molded plastic with a fluid well that contains numerous pads impregnated with specific chemical reagents. A drop of whole blood is applied to the center of the well and reacts with the chemicals in the pad to produce a specific color that is measured from the bottom of the card.26
Cell and platelet counts
Traditional cell-counting methods can be employed at the point of care for the analysis of WBCs, RBCs, and platelets. The Ichor Hematology Analyzer (Helena Laboratories, Beaumont, TX) performs a complete blood count along with platelet aggregation. Another option for cell quantitation and differentiation employs a buffy coat analysis method. Quantitative buffy coat analysis (QBC STAR, manufactured by QBC Diagnostics, Inc., Philipsburg, PA) involves centrifugation in specialized capillary tubes designed to expand the buffy coat layer. The components (platelets, mononuclear cells, and granulocytes) can be measured with the assistance of fluorescent dyes and a measuring device.27
Summary
Now that you have completed this chapter, read again the case studies at the beginning and respond to the questions presented.
Review questions
RBCs = 5.00 × 1012/L
HGB = 9 g/dL
HCT = 30%
|
MCV (fL) |
MCHC (g/dL) |
|
a. 30 |
18 |
|
b. 60 |
30 |
|
c. 65 |
33 |
|
d. 85 |
35 |
Observed reticulocyte count = 5.3%
HCT = 35%
Morphology—moderate polychromasia
Observed reticulocyte count = 6%
HCT = 30%
References
Additional resources